Sign in →

Test Code CA199 Carbohydrate Antigen 19-9 (Cancer Antigen 19-9)

Additional Codes

PathLab CA199
Epic ID LAB777

 

Specimen Requirements

Preferred Collection Container: 

Light Green (Lithium Heparin Plasma Separator Tube)
SST (Serum Separator)

 

Also Acceptable Collection Container: 

Green No Gel (Lithium Heparin)
Red (No Additive)
Plasma from Lavender tubes is acceptable but will not route in the Laboratory Information System (Sunquest).

 

Collection Information: 

Collect blood in a Light Green (Lithium Heparin Plasma Separator tube) or SST (Serum Separator).

 

Transport Information: 

Centrifuge and aliquot 0.6 mL of serum or plasma into a plastic transport vial. Send refrigerated.

 

Minimum Sample Volume: 

0.5 mL of serum or plasma

Stability/Transport

Temperature Stability Preferred Transport
Refrigerated 14 days X
Frozen 3 months  
Room Temp 5 days  

 

Performing Lab

Pathology Laboratory

Day(s) Performed

Daily

Report Available

2 hours

CPT Code Information

86301