Test Code CO2 Carbon Dioxide (CO2)
Additional Codes
| Epic ID | LAB55 |
Specimen Requirements
Preferred Collection Container: green top heparin tube, SST
Also Acceptable Collection Container: serum red top tube
Collection Information:
Collect 0.5 mL of Lithium heparinized plasma or serum.
Serum or plasma must be separated from cells within 2 hours of collection.
Underfilling of tubes may account for low CO2 results of up to 3 mmol/L.


Stability/Transport
| Temperature | Stability | Preferred Transport |
| Refrigerated | 7 days | X |
Tube must remain tightly stoppered.
Unacceptable Conditions
Opened sample tubes.
Performing Lab
Allen Hospital Laboratory
Marshalltown Laboratory
Day(s) Performed
Daily
Report Available
1 hour
Reference Ranges
22-29 mmol/L
Critical Value: <10 mmol/L or >40 mmol/L
CPT Code Information
82374