Sign in →

Test Code PLA2RM Phospholipase A2 Receptor, Monitoring, Enzyme-Linked Immunosorbent Assay, Serum

Additional Codes

Mayo PLA2M
Epic ID LAB6563

 

Specimen Type

Serum


Specimen Required


Supplies: Sarstedt Aliquot Tube, 5 mL (T914)

Collection Container/Tube:

Preferred: Serum gel

Acceptable: Red top

Submission Container/Tube: Plastic vial

Specimen Volume: 1 mL

Collection Instructions: Centrifuge and aliquot serum into plastic vial


Specimen Minimum Volume

0.5 mL

Specimen Stability Information

Specimen Type Temperature Time
Serum Refrigerated (preferred) 14 days
  Frozen  14 days
  Ambient  8 hours
UPH Waterloo Region Lab Test Catalog Note:

Transport refrigerated.

Reject Due To

Gross hemolysis Reject
Gross lipemia OK
Gross icterus OK
 

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Method Name

Enzyme-Linked Immunosorbent Assay (ELISA)

Useful For

Distinguishing primary from secondary membranous nephropathy

 

Monitoring patients with membranous nephropathy, over time, for trends in anti-phospholipase A2 receptor antibody levels

Day(s) Performed

Monday through Friday

Report Available

3 to 7 days

Reference Values

<14 RU/mL: Negative

14 to 19 RU/mL: Borderline

≥20 RU/mL: Positive

CPT Code Information

83520