Sign in →

Test Code VITE Vitamin E

Additional Codes

PathLab VITE
Epic ID LAB130

 

Specimen Requirements

Preferred Collection Container: 

Light Green (Lithium Heparin Plasma Separator Tube)
SST (Serum Separator)

 

Also Acceptable Collection Container: 

Green No Gel (Lithium Heparin)
Red (No Additive)

 

Collection Information: 

Collect blood in a Light Green (Lithium Heparin Plasma Separator Tube) or SST (Serum Separator) tube. Protect sample from light.

 

Transport Information: 

Centrifuge and aliquot 1.0 mL of serum/plasma into an amber plastic transport vial. Protect from light. Send refrigerated.

 

Minimum Sample Volume: 

0.5 mL of serum/plasma

Stability/Transport

Temperature Stability Preferred Transport
Refrigerated 7 days
Frozen 30 days  

PROTECT FROM LIGHT

Unacceptable Conditions

Not protected from light

Performing Lab

Pathology Laboratory

Methodology

LC-MS/MS (Liquid Chromatography-Tandem Mass Spectrometry)

Day(s) Performed

Twice weekly

Report Available

2-3 days

CPT Code Information

84446